Phantom-based training of ultrasound-guided breast biopsy in medical education: a randomized controlled trial comparing handheld and high-end ultrasound

Abstract

Background

Modern handheld ultrasound devices (HUDs) are attractive for teaching programs in undergraduate medical education due to their miniaturization and portability along at relatively low cost. They offer high-resolution imaging and are easy to use, opening up new possibilities for training of novices in ultrasound (US)-guided percutaneous procedures. The objective of this study was to investigate if guidance by HUD is on par with a cart-based high-end ultrasound scanner (HEUS) regarding performance times and success rates in video- and phantom-based training of novices in US-guided freehand breast biopsy.

Methods

32 medical students without any experience in performing US-guided percutaneous biopsies, who had previously completed a standardized diagnostic US training program, were randomized into either a HUD-group (n = 16) or a HEUS-group (n = 16). After a video training lecture participants performed US-guided biopsies of hypoechogenic and hyperechogenic target-lesions in a breast phantom using either a HUD or a HEUS. Performance times and success rates were primary outcomes. Participants were asked to complete a post-study questionnaire (Likert Scale and Raw NASA Workload Task Load Index) for subjective assessment of the operability and individually perceived workload of both US imaging tools and guidance-techniques as secondary outcomes.

Results

Biopsy success rates were slightly higher using the HUD (79.7%) in comparison to the HEUS (68.8%, p = 0.045). Median performance times were similar for the HUD (0.63 min, interquartile range IQR = 0.37–1.08 min) compared to the HEUS (0.60 min, IQR = 0.30–2.09 min, p = 0.751). Operability and the individually perceived workload were rated equal.

Conclusions

Percutaneous biopsy performed by novices using HUDs is feasible, performance times, success rates, operability and the individually perceived workload were on par with HEUS-guidance. HUDs can be used as cost-effective tools for percutaneous biopsy training purposes in medical education.

Peer Review reports

Introduction

Major advances in US technology have enabled the widespread use of handheld ultrasound devices (HUDs) for many applications in various fields of clinical medicine [1, 2]. There is expanding literature describing the use of HUDs for core Point-of-Care-Ultrasound (POCUS) applications, including eFAST (extended focused assessment with sonography for trauma), identification of first trimester intrauterine pregnancy and other gynecological/obstetric indications, focused cardiac and thoracic US as well as procedural guidance of vascular access [1, 3, 4]. The probes can be carried in a lab coat and WLAN-connected to smartphones and tablets. Besides their lightweight nature and flexibility, HUDs are characterized by ease of handling, making them attractive for medical education training programs. As the use of bedside POCUS as extension of physical examination becomes customary, gaining familiarity with sonographic imaging techniques in medical school is considered to be beneficial [5]. Peer-to-peer programs for undergraduate POCUS training have been established at many institutions and scientifically evaluated [6]. However, current data still suggest a lack of structured undergraduate training in European medical universities, which is why the European Federation for Ultrasound in Medicine and Biology (EFSUMB) continues to promote the introduction of US as an integrative part of the core curriculum of medical education [7].

The various HUDs available on the market show specific characteristics such as the provided probes and software, e.g. Lumify (Philips Healthcare, Netherlands) has been rated with the highest image quality and Vscan Air (General Electric Healthcare, USA) the best in terms of ease of use [8]. In a recent comparison of eight different brands of HUDs, Vscan Air scored highest both in terms of B-Mode image quality and in terms of clinical significance [9]. Despite the rapid increase in usability and image resolution, imaging based on HUDs is still a complex and operator-dependent modality requiring standardized training und supervised practicing. Scientific data regarding the use of handheld US training involving undergraduate students are limited and heterogeneous. It is unclear whether HUDs are suitable for the training of novices [10]. Moreover, data on undergraduate training in US-guided percutaneous biopsy procedures are limited [11,12,13], and to the best of our knowledge, studies on biopsy training involving HUDs are lacking. To what extent and in what way HUDs can be integrated into medical undergraduates’ education still need to be clarified. While structured teaching in diagnostic US imaging as well as peer-to-peer programs have been running for years at our university hospital, no standardized practical interventional US training has been offered to medical students until now. This study aimed at evaluating if performance and success rates of novices conducting US-guided, video- and phantom-based breast biopsies differ depending on the used image guidance device (HUD vs. HEUS).

Materials and methods

Study design and participants

This single-center prospective randomized study was conducted at the Department of Radiology at the University Medical Center Regensburg between April 2024 and August 2024 after institutional review board approval. As this educational study did not involve clinical interventions it was not prospectively registered (Clinical trial number: not applicable). Medical Center teaching staff of the School of Medicine invited 120 medical students who had passed the first state examination and had attended a standardized training in diagnostic US. Obligatory for inclusion was the completion of two consecutive ultrasound courses in a near-peer-teaching setting, addressing basical as well as advanced knowledge of performing standardized ultrasound examinations on various body parts, though not covering interventional US-techniques so far [14, 15]. 32 medical students participated as volunteers in the study and written informed consent was obtained before study inclusion. Participants were randomly assigned to one of the two groups, the HUD-group or HEUS-group, using an envelope randomization strategy (Fig. 1).

Fig. 1
figure 1

Diagrammatic representation of the study design

The time slots, for which the participants were scheduled, were assigned to the HUD-group or HEUS-group by drawing envelope-concealed numbered assignment slips from an opaque pouch.

At the scheduled time slot the participants viewed one of two prerecorded training videos (approx. 9 min) providing a general overview of how US-guided interventions are performed and showing the specific task the participants were to perform (video #1 for using LOGIQ E9, and video #2 for using Vscan Air CL; both US-devices by General Electric Healthcare, Chicago, USA). The structure of the breast phantom and the target lesions were presented in the video. For standardization purposes participants received training solely through video presentation. Each participant watched the assigned video alone in a room, pausing or repeating it in whole or in part was not allowed. They could familiarize themselves with the assigned US-devices, using the linear probe of LOGIQ E9 with a frequency range of 9-14 MHz or Vscan Air with a frequency range of 3–12 MHz, the transparent and opaque breast phantoms (Kyoto Kagaku, Kyoto, Japan; Fig. 2c) and the biopsy needle (Achieve 14G, 11.0 cm long, Merit Medical, Limburg, Germany) during the video lecture and up to 5 min afterwards. The HUD was connected via WLAN to an ipad (Apple Inc., Cupertino, USA) solely serving for this purpose (Fig. 2a).

Fig. 2
figure 2

Successful biopsy. Post-fire image using handheld ultrasound guidance in a. Needle depicted in hyperechogenic lesion (arrow). Core sample with blue colored material indicating successful biopsy in b. Opaque (left and transparent (right) breast phantom in c

Afterwards the participant scanned the breast phantom in B-Mode for identification of 4 target lesions: One 10 mm hypoechogenic and one 10 mm hyperechogenic lesion in the nipple-near part of the phantom, one 6 mm hyperechogenic lesion in the middle and one 10 mm hyperechogenic lesion in the deeper part of the phantom (Fig. 3). All participants had to biopsy the same 4 lesions in the same order starting with the 2 more superficial lying lesions and ending with the lesion in the deeper part of the phantom.

Fig. 3
figure 3

Target lesions. Hypoechogenic lesions in a and c, hyperechogenic lesions in b and d. Biopsy needle (arrows) pre-fire in c, post-fire in d. Hyperechogenic channels from previous biopsies (arrowheads)

Once a target lesion was identified and adjusted, the preloaded biopsy needle was freehand-inserted into the phantom by the participant under US-guidance either using the HUD or the HEUS. At this timepoint (contact of the needle tip with the surface of the phantom) a timer was started by a supervisor, who was present in the same room without interacting with the participant. Time measurement was finished as soon as a core biopsy was obtained by firing the biopsy needle (Fig. 2a). A biopsy was deemed successful if any colored lesion material was present in the core sample (Fig. 2b). Four phantoms were used for the study (2 in each group) and the phantoms were exchanged for a new one in the middle of the study after the eighth participant in both groups. It had been outlined in the video training how to differentiate between tracks by prior biopsies and the inserted biopsy needle.

The procedure was repeated two times for each of the four lesions (in total 8 times per participant). Following the biopsy procedures, all participants were asked to complete a post-procedural questionnaire (Supplemental Fig. 1). For evaluating and assessing the participants` subjective impressions in the trial they had to record their personal experience in three categories. The Raw NASA Workload Task Load Index (TLX), a subjective, multidimensional assessment tool was used to rate the participants` perceived workload [16, 17]. The questionnaire`s design was chosen to be simple to facilitate the students´ assessment and minimize interobserver variability between the evaluators. Moreover, the standardized protocol including video-based training with as little interaction between observers and participants as possible and the choice of objective and simple parameters as primary outcomes were important factors addressing interobserver variability. For these reasons, the interobserver variability was not quantified.

Primary outcomes

Primary outcomes of the study were the rate of positive biopsies per student and the skin-to-fire-time for each lesion.

Secondary outcomes

The subjective assessment of the operability and safety of US imaging and biopsy guidance devices and individually perceived workload were defined as secondary outcomes. Directly after the biopsies were taken, the participants had to record their personal impressions in three categories with a 6-point Likert scale (0–5; 0 = absolutely comfortable/safe and not difficult at all, 5 = not at all comfortable resp. safe and very difficult): (1) Comfort (“I felt the imaging system used was comfortable for me”), (2) Safety (“I felt safe performing the biopsy while using this ultrasound imaging system“, regarding orientation, needle firing and tissue asservation), (3) Difficulty (“Overall difficulty of the task“). The assessment of the mental workload was performed with the Raw NASA Task Load Index (NASA TLX). Mental, physical and temporal demand, performance, effort and frustration of the participants were assessed on a 20-point Likert scale (1–20; 1 = very low, 20 = very high).

Statistical analysis

Frequencies are presented as absolute numbers and percentages. Continuous data are presented as median values and interquartile ranges (IQRs). Differences between the groups were analyzed using the Pearson chi-square test of independence for dichotomous parameters and the Wilcoxon-Mann-Whitney U test for continuous data. A p-value < 0.05 was considered statistically significant. All analyses were performed using SPSS version 29.0.1.0 (IBM, NY, USA).

Results

Overall biopsy times were similar overall and for every single target lesion in the HUD-group (0.63 min, IQR = 0.37–1.08 min) and the HEUS-group (0.60 min, IQR = 0.30–2.09 min) with no statistically significant difference between the two devices (p = 0.751) (Fig. 4). For both devices median biopsy times for the last deep lying lesion were slightly longer than for the first superficial lying lesion: 0.70 min, IQR 0.44–1.71 (HUS) and 0.94 min, IQR 0.63–1.22 min (HUD) versus 0.54 min, IQR = 0.24–2.17 (HUS) and 0.77 min, IQR 0.44–1.19 min (HUD). Biopsies guided with the HUD were significantly more often successful in comparison to guidance with HEUS (79.9% versus 68.8%, p = 0.045).

Fig. 4
figure 4

Boxplots showing intervention times for the devices used

With regard to the secondary outcomes comfort and safety, the HUD achieved slightly better ratings with no significant difference. The evaluation of mental workload assessment showed no significant differences between the ratings given by the two groups concerning all items (Table 1).

Table 1 Subjective assessment of the impressions regarding comfort, safety and difficulty and personal workload

Discussion

The feasibility of US-guided procedures carried out by medical students has been examined mostly with US-guided vascular access, much rarer with percutaneous biopsies [18]. Limited resources on staff and budget are opposed to a growing need for US teaching facilities, as pointed out by Schmidt et al., who assessed the effectiveness of a US breast training program including percutaneous biopsies on knowledge and skills of undergraduate medical students, concluding that medical education can be enhanced by teaching breast US skills [19].

The purpose of our initial feasibility study focused on the applicability of HUDs in novices regarding US-guided procedures and a more resource-saving setting such as training by video lectures. Our results suggest a non-inferiority of the used HUD compared to the HEUS for US-guided percutaneous breast phantom biopsy training. Biopsies guided with the HUD were slightly more successful than performed with the HEUS. We assume this small difference is due to technical advantages of the HUD such as probe flexibility or the individually adjustable positioning options of the used tablet, but random variability cannot be completely excluded due to the small study cohort. Certainly, a follow-up cross over-study with focus on the technical differences between the devices and including more participants would be beneficial in addressing this question.

No significant difference between the devices was found regarding overall biopsy times and the subjective assessment of comfort and safety in procedure guidance. Even though older data described a higher acceptance of HEUS amongst students compared to HUD for POCUS training [20], modern HUDs are used more frequently nowadays [21]. In our experimental setting the medical students who had passed their standardized diagnostic US training using HEUS rated their first experiences with HUD similar regarding the individually perceived workload, comfort and safety indicating a high acceptance for this new technique.

The question arises whether the use of HUDs could support peer-to-peer or near-peer education for undergraduate student training. The utilization of near-peers in the practice of ultrasound-guided procedures had been described as effective to develop sonography skills while reinforcing anatomy [22]. Medical education is one of the most interesting areas of application for HUD and has been described as a driving force for innovation of mobile US technique [4, 23]. Besides the ease of handling, cost-effectiveness is a major advantage compared to the usually more expensive cart-based systems, making it an attractive tool for students. HUDs have the potential to increase the availability of US training in medical education and to facilitate early exposure to US-guided procedure skills training by replacing the more expensive standard HEUS. Insufficient teaching on interventional radiology techniques in medical schools has been described as one reason for the workforce shortages in interventional radiology [24] and standardized curricula with a greater focus on interventional radiology have been suggested [25]. Lower costs of HUDs could ensure higher availability of devices used for hands-on interventional skills training. Given their ease of handling, additional training on the usage of HUDs specifically would probably not be required, but further research on the influence of HUDs on US training is needed. Future research should also focus on developing objective assessment criteria to measure the proficiency of ultrasound skills as stated by Recker et al. [26]. The long-term impacts of undergraduate handheld US training should be evaluated and evidence-based guidelines for targeted education need to be implemented. Methodological requirements for teaching studies have been recommended in order to establish future guidelines on US training in medical education [27].

Moreover, the comparatively low costs enable the use of portable devices in low-to middle income countries or other resource-limited settings, opening up new possibilities not only for general POCUS applications, but also for US-guided procedure training of undergraduate and graduate students [28,29,30]. Innovative low-cost solutions such as HUDs offer a safe, simple and sustainable solution toward building capacity for cancer control in low- to middle-income countries that at the moment mostly rely on blind or surgical biopsies for cancer diagnosis, as pointed out by Hricak et al. [31].

There are growing data on the use of deep learning algorithms to facilitate POCUS operations in resource-constrained settings [32]. Artificial intelligence may provide real-time coaching of image generation and interpretation in the future. In general, widespread use of artificial intelligence is expected to enhance the feasibility of prospective POCUS applications.

A prerequisite for the unrestricted use of HUDs in clinical routine is a secure image documentation. According to our experience, data transfer to and storage in the picture archiving and communication system as the traditional solution for image archiving are central issues that have not yet been solved satisfactorily. Cloud-based solutions for storage of de-identified data might be beneficial, especially in out-of-hospital situations, but still require regulatory challenges. While data security should be considered at all times when patients are involved, this topic can advantageously be neglected in the context of phantom- or simulator-based student training. In the setting of US skills training no requirements and legal regulations exist regarding image archiving.

The phantoms used for this study reproduced the resistance and softness of breast tissue to a satisfactory degree. A minor limitation was the presence of hyperechogenic biopsy channels, which made depicting the needle tip during the biopsy slightly more difficult. However, this limitation was consistent across both groups, as two phantoms were used per group and were replaced with new ones midway through the study, after the eighth participant in each group.

A problem we had to face was that the HUD`s scanning time is limited by the battery [33]. At least two HUDs were needed to carry out the study, for one device usually had to be charged on site. This should be taken into account especially when several trainees are participating. It is to be expected that the restriction of scanning times can be reduced by higher battery capacities in the future. The limited number of participants has to be stated as a major limitation. This is due to the strict inclusion criteria which were defined to ensure the formation of two highly homogeneous study groups with the same experience level regarding US.

Conclusions

We believe interventional US training should be part of the curriculum in medical education. Costs incurred could be significantly reduced and availability increased through the use of HUDs instead of cart-based machines. While this study showed a non-inferiority of HUDs regarding biopsy success rates, intervention times and the individually perceived comfort, safety and workload of the medical students in phantom-based percutaneous biopsy training, further research is needed on long-term skill-retention, standardized training protocols and the value of AI-assisted guidance.

Data availability

The data and materials in this paper are available from the corresponding author on request.

Abbreviations

HEUS:

High-end ultrasound

HUD:

Handheld ultrasound device

IQR:

Interquartile range

POCUS:

Point-of-Care-Ultrasound

US:

Ultrasound

References

  1. Nielsen MB, Cantisani V, Sidhu PS, Badea R, Batko T, Carlsen J, et al. Die Anwendung handgeführter Ultraschallgeräte– Ein EFSUMB positionspapier. [The Use of Handheld Ultrasound Devices - An EFSUMB Position Paper]. Ultraschall Med. 2019;40:30–9. https://doiorg.publicaciones.saludcastillayleon.es/10.1055/a-0783-2303.

    Article  Google Scholar 

  2. Rykkje A, Carlsen JF, Nielsen MB. Hand-Held ultrasound devices compared with High-End ultrasound systems: A systematic review. Diagnostics. 2019;9:61. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/diagnostics9020061.

    Article  Google Scholar 

  3. Thavanathan RS, Woo MY, Hall G. The future is in your hands - Handheld ultrasound in the emergency department. CJEM. 2020;22:742–4. https://doiorg.publicaciones.saludcastillayleon.es/10.1017/cem.2020.449.

    Article  Google Scholar 

  4. Malik AN, Rowland J, Haber BD, Thom S, Jackson B, Volk B, Ehrman RR. The use of handheld ultrasound devices in emergency medicine. Curr Emerg Hosp Med Rep. 2021;9:73–81. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s40138-021-00229-6.

    Article  Google Scholar 

  5. Hand-held ultrasound scanners in medical education: a systematic review. 2017. J Cardiovasc Ultrasound. 2017;25(3):75–83. https://doiorg.publicaciones.saludcastillayleon.es/10.4250/jcu.2017.25.3.75

  6. Sachs A. Sono4You - peer-assisted ultrasound tutorials for undergraduate medical students. Eur Congress Radiol. 2013. https://doiorg.publicaciones.saludcastillayleon.es/10.1594/ecr2013/C-2546.

    Article  Google Scholar 

  7. Prosch H, Radzina M, Dietrich CF, Nielsen MB, Baumann S, Ewertsen C, et al. Ultrasound curricula of student education in Europe: summary of the experience. Ultrasound Int Open. 2020;6:E25–33. https://doiorg.publicaciones.saludcastillayleon.es/10.1055/a-1183-3009.

    Article  Google Scholar 

  8. Le M-PT, Voigt L, Nathanson R, Maw AM, Johnson G, Dancel R, et al. Comparison of four handheld point-of-care ultrasound devices by expert users. Ultrasound J. 2022;14:27. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13089-022-00274-6.

    Article  Google Scholar 

  9. Merkel D, Züllich TF, Schneider C, Yousefzada M, Beer D, Ludwig M et al. Prospective Comparison of Handheld Ultrasound Devices from Different Manufacturers with Respect to B-Scan Quality and Clinical Significance for Various Abdominal Sonography Questions. Diagnostics. 2023. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/diagnostics13243622

  10. Haji-Hassan M, Capraș R-D, Bolboacă SD. Efficacy of handheld ultrasound in medical education: A comprehensive systematic review and narrative analysis. Diagnostics. 2023;13:3665. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/diagnostics13243665.

    Article  Google Scholar 

  11. Schmidt G, Gerlinger C, Endrikat J, Gabriel L, Müller C, Baus S, et al. Teaching breast ultrasound skills including core-needle biopsies on a Phantom enhances undergraduate student’s knowledge and learning satisfaction. Arch Gynecol Obstet. 2021;304:197–202. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s00404-021-06016-8.

    Article  Google Scholar 

  12. Wampler K, Bartenhagen L, Sayles H, Michael K. Lightly embalmed cadavers in sonography education: A novel approach to improving students’ perceived competence in renal biopsy procedures. J Diagn Med Sonography. 2020;36:3–10. https://doiorg.publicaciones.saludcastillayleon.es/10.1177/8756479319875451.

    Article  Google Scholar 

  13. Nayahangan LJ, Dietrich CF, Nielsen MB. Simulationsbasiertes training Im Ultraschall– wo Stehen wir? [Simulation-based training in ultrasound - where are we now?]. Ultraschall Med. 2021;42:240–4. https://doiorg.publicaciones.saludcastillayleon.es/10.1055/a-1352-5944.

    Article  Google Scholar 

  14. Wolf R, Geuthel N, Gnatzy F, Rotzoll D. Undergraduate ultrasound education at German-speaking medical faculties: a survey. GMS J Med Educ. 2019;36:Doc34. https://doiorg.publicaciones.saludcastillayleon.es/10.3205/zma001242.

    Article  Google Scholar 

  15. https://. fsmed-regensburg.de/arbeitsgruppen/ag-sono4you. Accessed 28 January 2025.

  16. Buia A, Stockhausen F, Filmann N, Hanisch E. 2D vs. 3D imaging in laparoscopic surgery-results of a prospective randomized trial. Langenbecks Arch Surg. 2017;402:1241–53. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s00423-017-1629-y.

    Article  Google Scholar 

  17. Hart SG. Nasa-Task Load Index (NASA-TLX); 20 Years Later. 2006. https://doiorg.publicaciones.saludcastillayleon.es/10.1177/154193120605000909.

  18. Tarique U, Tang B, Singh M, Kulasegaram KM, Ailon J. Ultrasound curricula in undergraduate medical education: A scoping review. J Ultrasound Med. 2018;37:69–82. https://doiorg.publicaciones.saludcastillayleon.es/10.1002/jum.14333.

    Article  Google Scholar 

  19. Schmidt G, Gerlinger C, Endrikat J, et al. Teaching breast ultrasound skills including core-needle biopsies on a Phantom enhances undergraduate student’s knowledge and learning satisfaction. Arch Gynecol Obstet. 2021;304(1):197–202. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s00404-021-06016-8.

    Article  Google Scholar 

  20. Gogalniceanu P, Sheena Y, Kashef E, Purkayastha S, Darzi A, Paraskeva P. Is basic emergency ultrasound training feasible as part of standard undergraduate medical education? J Surg Educ. 2010;67:152–6. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.jsurg.2010.02.008.

    Article  Google Scholar 

  21. McCormick E, Flanagan B, Johnson CD, Sweeney EM. Ultrasound skills teaching in UK medical education: A systematic review. Clin Teach. 2023;20:e13635. https://doiorg.publicaciones.saludcastillayleon.es/10.1111/tct.13635.

    Article  Google Scholar 

  22. Martin R, Lau HA, Morrison R, Bhargava P, Deiling K. The rising tide of Point-of-Care ultrasound (POCUS) in medical education: an essential skillset for undergraduate and graduate medical education. Curr Probl Diagn Radiol. 2023;52(6):482–4. https://doiorg.publicaciones.saludcastillayleon.es/10.1067/j.cpradiol.2023.06.003.

    Article  Google Scholar 

  23. Slader M, Young H, Barker M, Prentice K, Bogaard K, Yuan C, et al. A comparison of handheld and standard ultrasound in Swiss medical students. World J Emerg Med. 2022;13:85–90. https://doiorg.publicaciones.saludcastillayleon.es/10.5847/wjem.j.1920-8642.2022.017.

    Article  Google Scholar 

  24. Osman FH, Koe JSE, Lau ESW, Nagaraj D, Ng HH-L, Ong LA, et al. Evaluation of learning and exposure in the undergraduate interventional radiology curriculum (ELIXIR). Clin Radiol. 2023;78:e773–81. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.crad.2023.07.001.

    Article  Google Scholar 

  25. Foo M, Maingard J, Phan K, Lim R, Chandra RV, Lee MJ, et al. Australian students’ perspective on interventional radiology education: A prospective cross-institutional study. J Med Imaging Radiat Oncol. 2018;62:758–63. https://doiorg.publicaciones.saludcastillayleon.es/10.1111/1754-9485.12764.

    Article  Google Scholar 

  26. Recker F, Neubauer R, Dong Y, et al. Exploring the dynamics of ultrasound training in medical education: current trends, debates, and approaches to didactics and hands-on learning. BMC Med Educ. 2024;24:1311. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12909-024-06092-9.

    Article  Google Scholar 

  27. Neubauer R, Bauer CJ, Dietrich CF, Strizek B, Schäfer VS, Recker F. Evidence-based ultrasound Education?– A systematic literature review of undergraduate ultrasound training studies. Ultrasound Int Open. 2024. https://doiorg.publicaciones.saludcastillayleon.es/10.1055/a-2275-0702.

    Article  Google Scholar 

  28. Munyaneza R, Martin AN, Riviello R, Abahuje E, Murthy SS, Ntirenganya F. Teaching surgical skills in a resource-limited setting: comparing massed versus distributed practice in an ultrasound-guided breast biopsy simulator. East Cent Afr J Surg. 2019;23:89–94. https://doiorg.publicaciones.saludcastillayleon.es/10.4314/ecajs.v23i3.1.

    Article  Google Scholar 

  29. Hey MT, Alayande BT, Masimbi O, Shimelash N, Forbes C, Twizeyimana J, et al. Developing a surgical simulation curriculum for the Rwandan context. J Surg Educ. 2023;80:1268–76. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.jsurg.2023.06.007.

    Article  Google Scholar 

  30. Becker DM, Tafoya CA, Becker SL, Kruger GH, Tafoya MJ, Becker TK. The use of portable ultrasound devices in low- and middle-income countries: a systematic review of the literature. Trop Med Int Health. 2016;21:294–311. https://doiorg.publicaciones.saludcastillayleon.es/10.1111/tmi.12657.

    Article  Google Scholar 

  31. Hricak H, Abdel-Wahab M, Atun R, et al. Medical imaging and nuclear medicine: a lancet oncology commission. Lancet Oncol. 2021;22(4):e136–72. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/S1470-2045(20)30751-8.

    Article  Google Scholar 

  32. Kim S, Fischetti C, Guy M, Hsu E, Fox J, Young SD. Artificial intelligence (AI) applications for point of care ultrasound (POCUS) in Low-Resource settings: A scoping review. Diagnostics. 2024. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/diagnostics14151669.

    Article  Google Scholar 

  33. Wilkinson JN, Saxhaug LM. Handheld ultrasound in training - The future is getting smaller! J Intensive Care Soc. 2021;22:220–9. https://doiorg.publicaciones.saludcastillayleon.es/10.1177/1751143720914216.

    Article  Google Scholar 

Download references

Acknowledgements

We sincerely thank the volunteers for their participation. We would like to express our gratitude to the medical students Julia Feicht and Lennart Philipp, leading tutors of Sono4You at our faculty, for supporting the study.

Funding

Open Access funding enabled and organized by Projekt DEAL.

No funding was received for this study.

Author information

Authors and Affiliations

Authors

Contributions

BG and SH contributed to the conception, design of the work, acquisition and interpretation of data, and drafted the work. FZ and MB contributed to the statistical analysis and interpretation of data. MB prepared figs. 1 and 4. MA, AG, EMJ and CS contributed to the conception of the work and revised the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Barbara Greiner.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Research Ethics Committee of the University of Regensburg (report number: 24-3745-104). Written informed consent was obtained from all participating medical students. All respondents participated in this study freely and with consent. The study was performed in accordance with relevant guidelines and regulations and in compliance with the Helsinki Declaration.

Consent for publication

All participating medical students had given their written consent for data publication.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Greiner, B., Akers, M., Zeman, F. et al. Phantom-based training of ultrasound-guided breast biopsy in medical education: a randomized controlled trial comparing handheld and high-end ultrasound. BMC Med Educ 25, 551 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12909-025-07163-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12909-025-07163-1